Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Nitinol Staple Recalled by TriMed Inc. Due to During a retrospective review of sterilization records, TriMed...

Date: April 13, 2018
Company: TriMed Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact TriMed Inc. directly.

Affected Products

Nitinol Staple, Elastic Staple Small-Asymmetrical, Sterile R, Model #: a) ES-10x13x15; b) ES-10x16x19; c) ES-10x17x19; d) ES-12x13x15; e) ES-12x15x17; f) ES-12x16x19; g) ES-15x13x16

Quantity: 706

Why Was This Recalled?

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Where Was This Sold?

This product was distributed to 14 states: CA, FL, IL, KS, MN, MS, MO, NV, NJ, NC, PA, TX, UT, WA

Affected (14 states)Not affected

About TriMed Inc.

TriMed Inc. has 18 total recalls tracked by RecallDetector.

Related Recalls