Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

7 Hole VL Gridlock Fibula Plate Recalled by Trilliant Surgical Ltd. Due to The affected parts subject to the recall are...

Date: August 6, 2015
Company: Trilliant Surgical Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Trilliant Surgical Ltd. directly.

Affected Products

7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY; 9 Hole VL Gridlock Fibula Plate, Part # 300-60-002 NON-STERILE SINGLE USE ONLY; 11 Hole VL Gridlock Fibula Plate, Part # 300-60-003 NON-STERILE SINGLE USE ONLY. For use in trauma and reconstructive procedures in small bones.

Quantity: 28 units

Why Was This Recalled?

The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.

Where Was This Sold?

This product was distributed to 7 states: CA, FL, MN, NJ, NM, OH, WI

Affected (7 states)Not affected

About Trilliant Surgical Ltd.

Trilliant Surgical Ltd. has 10 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report