Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Concave Reamer Head Recalled by Trilliant Surgical Ltd. Due to The cannulation of the Concave and Convex Reamer...

Date: July 20, 2015
Company: Trilliant Surgical Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Trilliant Surgical Ltd. directly.

Affected Products

Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.

Quantity: 638 units

Why Was This Recalled?

The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.

Where Was This Sold?

This product was distributed nationwide across all 50 states.