Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

2.4 Recalled by Trilliant Surgical Ltd. Due to Gridlock screw driver bits for the Gridlock Plating...

Date: July 31, 2015
Company: Trilliant Surgical Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Trilliant Surgical Ltd. directly.

Affected Products

2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.

Quantity: 108 units

Why Was This Recalled?

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Where Was This Sold?

This product was distributed to 19 states: AZ, AR, CA, FL, GA, IL, IN, IA, MN, MO, NJ, NY, OH, OK, SC, SD, TX, VA, WI

Affected (19 states)Not affected

About Trilliant Surgical Ltd.

Trilliant Surgical Ltd. has 10 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report