Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: Recalled by Trilliant Surgical, LLC Due to A screwdriver used to facilitate implant insertion and...

Date: March 14, 2019
Company: Trilliant Surgical, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Trilliant Surgical, LLC directly.

Affected Products

Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.

Quantity: 20 units

Why Was This Recalled?

A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Trilliant Surgical, LLC

Trilliant Surgical, LLC has 13 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report