Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Jasper Vektor Recalled by TP Orthodontics, Inc. Due to The firm has received 106 complaints of the...

Date: May 1, 2018
Company: TP Orthodontics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact TP Orthodontics, Inc. directly.

Affected Products

Jasper Vektor, Class II Correction Appliance, Part No. 610-536 used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops.

Quantity: 5

Why Was This Recalled?

The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient.

Where Was This Sold?

This product was distributed to 20 states: AL, CA, CO, FL, HI, ID, IL, ME, MA, MI, NV, NJ, NY, OH, OR, PA, SC, TX, UT, VA

Affected (20 states)Not affected

About TP Orthodontics, Inc.

TP Orthodontics, Inc. has 13 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report