Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AIA Analyzer Pipette Tips used on Tosoh AIA-2000 Recalled by Tosoh Smd Inc Due to Visual inspection of the pipette tips indicated that...

Date: May 15, 2013
Company: Tosoh Smd Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Tosoh Smd Inc directly.

Affected Products

AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyzer Tips are disposable pipette tips packaged in bulk quantities and used on Tosoh Bioscience AIA Analyzers for lab analysis of various immunoassays.

Quantity: 11,309 bags

Why Was This Recalled?

Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers.

Where Was This Sold?

This product was distributed to 44 states: AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY

Affected (44 states)Not affected

About Tosoh Smd Inc

Tosoh Smd Inc has 3 total recalls tracked by RecallDetector.

Related Recalls