Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tosoh¿Sorter 9tray Recalled by Tosoh Bioscience Inc Due to This recall has been initiated due to a...

Date: June 11, 2017
Company: Tosoh Bioscience Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Tosoh Bioscience Inc directly.

Affected Products

Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.

Quantity: 565 total devices U.S.

Why Was This Recalled?

This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Tosoh Bioscience Inc

Tosoh Bioscience Inc has 68 total recalls tracked by RecallDetector.