Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tosoh¿Automated Enzyme Immunoassay Analyzer AIA-900 Recalled by Tosoh Bioscience Inc Due to This recall has been initiated due to a...

Date: June 11, 2017
Company: Tosoh Bioscience Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Tosoh Bioscience Inc directly.

Affected Products

Tosoh¿Automated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930. Product Usage: The AIA-900 Automated Enzyme Immunoassay System is an automated analyzer for in vitro diagnostic use that measures a variety of substances in serum, plasma, and or urine using the AIA-PACK reagent series.

Quantity: 565 total devices U.S.

Why Was This Recalled?

This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Tosoh Bioscience Inc

Tosoh Bioscience Inc has 68 total recalls tracked by RecallDetector.