Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing. Recalled by Tosoh Bioscience, Inc. Due to The device defect associated with this recall may...

Date: November 7, 2016
Company: Tosoh Bioscience, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Tosoh Bioscience, Inc. directly.

Affected Products

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.

Quantity: 682 analyzers

Why Was This Recalled?

The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.