Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for Recalled by Tosoh Bioscience Inc Due to Decreased stability for five (5) lots of Calibrator...

Date: August 7, 2020
Company: Tosoh Bioscience Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Tosoh Bioscience Inc directly.

Affected Products

Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for BHCG Part number 020361 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay

Quantity: 289 boxes

Why Was This Recalled?

Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.

Where Was This Sold?

This product was distributed nationwide across all 50 states.