Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

TSX-301C Aquilion One (Vision Edition) Computed Tomography X-Ray System. Recalled by Toshiba American Medical Systems Inc Due to Toshiba America Medical Systems, Inc. initiated this recall...

Date: December 11, 2013
Company: Toshiba American Medical Systems Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Toshiba American Medical Systems Inc directly.

Affected Products

TSX-301C Aquilion One (Vision Edition) Computed Tomography X-Ray System.

Quantity: 8 units

Why Was This Recalled?

Toshiba America Medical Systems, Inc. initiated this recall due to the failure of the parts used in the X-ray high-voltage generator. In some cases, X-ray exposure cannot be performed.

Where Was This Sold?

This product was distributed to 8 states: CA, IN, MD, MN, NY, OH, WV, WI

Affected (8 states)Not affected

About Toshiba American Medical Systems Inc

Toshiba American Medical Systems Inc has 56 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report