Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that Recalled by Toshiba American Medical Systems Inc Due to Although the Flat Panel Detector (FPD) touch sensor...

Date: May 12, 2014
Company: Toshiba American Medical Systems Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Toshiba American Medical Systems Inc directly.

Affected Products

Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.

Quantity: 2

Why Was This Recalled?

Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.

Where Was This Sold?

This product was distributed nationwide across all 50 states.