Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

INFX-8000C Fluoroscopic X-Ray Systems Recalled by Toshiba American Medical Systems Inc Due to During an examination a device error occurred and...

Date: December 29, 2017
Company: Toshiba American Medical Systems Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Toshiba American Medical Systems Inc directly.

Affected Products

INFX-8000C Fluoroscopic X-Ray Systems

Quantity: 19 units

Why Was This Recalled?

During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.

Where Was This Sold?

United States