Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The Recalled by Toshiba American Medical Systems Inc Due to Potential problem with the cardiac function analysis software...

Date: November 5, 2014
Company: Toshiba American Medical Systems Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Toshiba American Medical Systems Inc directly.

Affected Products

Cardiac Function Analysis Software Product Data No. MPDCT0301EAD CSCF-003A The software is applicable to the following CT systems: Aquilion ONE TSX-301C X-ray CT system with a 320-row 0.5 mm detector is provided with a 7.5-MHU large-capacity X-ray tube unit, and permits scanning over large areas at the high-speed of 0.35 s per rotation. The system features are; 160-mm-wide area detector, station, gantry and patient couch operation controls, image reconstruction, high-speed volume scan, high-quality images, selectable image slice thickness, dual-monitor system, improved image diagnosis functions, high-speed volume data workflow, exposure reduction, ECG-gated scan and reconstruction, and SUREFluoro. The CT cardiac function analysis software is used to analyze cardiac function using data obtained by performing cardiac MPR processing for ECG-gated scan data according to the cardiac axis. Aquilion ONE is a dynamic volume CT system that supports whole-body scanning. This 320-detector row system generates 640 slices per rotation using the coneXact reconstruction algorithm. In addition, the high-speed rotation mechanism and the fast reconstruction unit of the system allow the rapid image acquisition to further improve throughput in CT examinations.

Quantity: 1 unit within the U.S. only

Why Was This Recalled?

Potential problem with the cardiac function analysis software (CFA). Incorrect analysis results may be displayed in the Function Parameters for the Entire Heart displayed as analysis results of the CFA and in a Left-Ventricular Volume Curve generated based on some of those parameters.

Where Was This Sold?

This product was distributed nationwide across all 50 states.