Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Aquilion Prime TSX-303A Aquilion PRIME is a multislice helical CT Recalled by Toshiba American Medical Systems Inc Due to Toshiba American Medical Systems (TAMS) is recalling the...

Date: October 25, 2016
Company: Toshiba American Medical Systems Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Toshiba American Medical Systems Inc directly.

Affected Products

Aquilion Prime TSX-303A Aquilion PRIME is a multislice helical CT system with an 40-row detector. High-speed rotation allows rapid data acquisition and shortens scan times, while the fast reconstruction unit further improves throughput, reducing the time required for diagnosis.

Quantity: 291 units

Why Was This Recalled?

Toshiba American Medical Systems (TAMS) is recalling the Aquilion CT system because it has been reported that the cooling oil in the x-ray tube heat exchanger may leak during system operation.