Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a Recalled by Topcon Medical Systems, Inc. Due to When a patient is selected in the patient...

Date: March 6, 2019
Company: Topcon Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Topcon Medical Systems, Inc. directly.

Affected Products

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

Quantity: 450 units

Why Was This Recalled?

When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.

Where Was This Sold?

This product was distributed nationwide across all 50 states.