Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PASCAL and PASCAL Streamline (532 nm.577 nm) Recalled by Topcon Medical Laser Systems, Inc Due to When using the "Favorites" function for saved settings,...

Date: March 27, 2012
Company: Topcon Medical Laser Systems, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Topcon Medical Laser Systems, Inc directly.

Affected Products

PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye.

Quantity: 7 US, 33 foreign

Why Was This Recalled?

When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.

Where Was This Sold?

Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.