Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Opthalmic: PASCAL Synthesis is an ophthalmic scanning laser system. The Recalled by Topcon Medical Laser Systems, Inc Due to Topcon discovered an anomaly with titration, only when...

Date: July 15, 2016
Company: Topcon Medical Laser Systems, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Topcon Medical Laser Systems, Inc directly.

Affected Products

Opthalmic: PASCAL Synthesis is an ophthalmic scanning laser system. The system can perform single shot photocoagulation as is performed conventionally.

Quantity: 205 units (21 in US, 184 OUS)

Why Was This Recalled?

Topcon discovered an anomaly with titration, only when EpM is turned on during the titration process. This event may occur if the user turns on EpM, turns off the Landmark (LM) function (with any EpM supported pattern) and then enters titration mode.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Topcon Medical Laser Systems, Inc

Topcon Medical Laser Systems, Inc has 2 total recalls tracked by RecallDetector.

Related Recalls