Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Binding Site Human IgM Kit (for SPAPLUS) Recalled by The Binding Site Group, Ltd. Due to Firm is offering advice to customers experiencing challenges...

Date: November 6, 2018
Company: The Binding Site Group, Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact The Binding Site Group, Ltd. directly.

Affected Products

Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgM antiserum, six calibrators and two controls. The calibrators and controls are based on stabilized human serum. All components contain sodium azide as a preservative and are provided in ready-to-use liquid format.

Quantity: 60

Why Was This Recalled?

Firm is offering advice to customers experiencing challenges when validating the calibration curve of the IgM kits. This is indicated by kit control values reporting outside the specified ranges.

Where Was This Sold?

This product was distributed to 10 states: CA, FL, IL, IN, NE, NY, PA, SD, VA, WA

Affected (10 states)Not affected

About The Binding Site Group, Ltd.

The Binding Site Group, Ltd. has 28 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report