Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Synthes Small Battery Drive (SBD) Oscillating Saw Attachment Part Number: Recalled by The Anspach Effort, Inc. Due to The potential exists for the Oscillating Saw Attachment...

Date: March 24, 2015
Company: The Anspach Effort, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact The Anspach Effort, Inc. directly.

Affected Products

Synthes Small Battery Drive (SBD) Oscillating Saw Attachment Part Number: 532.021 for orthopedic surgery.

Quantity: 12,085 pieces

Why Was This Recalled?

The potential exists for the Oscillating Saw Attachment to disengage from the Small Battery Drive handpiece. If the attachment disengages from the handpiece intra-operatively, there is a potential for injury to the patient and/or user.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About The Anspach Effort, Inc.

The Anspach Effort, Inc. has 261 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report