Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The Recalled by The Anspach Effort, Inc. Due to The RPM of the Motor Handpiece may be...

Date: June 24, 2011
Company: The Anspach Effort, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact The Anspach Effort, Inc. directly.

Affected Products

***REF SC2100***eMax 2 Plus Console***SN G 10307653002*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.

Quantity: 61 units

Why Was This Recalled?

The RPM of the Motor Handpiece may be less than 80,000 RPM.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About The Anspach Effort, Inc.

The Anspach Effort, Inc. has 261 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report