Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The Sarns Centrifugal Pump is a single use device sterilized Recalled by Terumo Cardiovascular Systems Corporation Due to Terumo Cardiovascular Systems Corporation has received complaints indicating...

Date: May 27, 2015
Company: Terumo Cardiovascular Systems Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Terumo Cardiovascular Systems Corporation directly.

Affected Products

The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single sterile in cartons of 8 units, or may be provided in bulk, non-sterile to another Terumo location for inclusion in Cardiovascular Procedure Kits. Cardiovascular Procedure kits are packaged individually.

Quantity: 579 units have entered into the distribution chain

Why Was This Recalled?

Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.

Where Was This Sold?

Terumo CVS has identified 33 domestic consignees and 0 international consignee that are affected by this recall

About Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Systems Corporation has 204 total recalls tracked by RecallDetector.