Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tested AC Control BD Recalled by Terumo Cardiovascular Systems Corporation Due to Based on a retrospective review of quality data,...

Date: October 16, 2012
Company: Terumo Cardiovascular Systems Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Terumo Cardiovascular Systems Corporation directly.

Affected Products

Tested AC Control BD, 120V, X2 The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Quantity: 15 units

Why Was This Recalled?

Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Systems Corporation has 204 total recalls tracked by RecallDetector.