Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation Recalled by Terumo Cardiovascular Systems Corporation Due to Inaccurate readings. Software contains unused (turned off) accuracy...

Date: November 1, 2010
Company: Terumo Cardiovascular Systems Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Terumo Cardiovascular Systems Corporation directly.

Affected Products

Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.

Quantity: 622 units

Why Was This Recalled?

Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.

Where Was This Sold?

Worldwide Distribution.

About Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Systems Corporation has 204 total recalls tracked by RecallDetector.