Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Terumo Advanced Perfusion System 1 Flow Module indicated for use Recalled by Terumo Cardiovascular Systems Corporation Due to Terumo Cardiovascular Systems (CVS) received one report of...

Date: September 20, 2012
Company: Terumo Cardiovascular Systems Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Terumo Cardiovascular Systems Corporation directly.

Affected Products

Terumo Advanced Perfusion System 1 Flow Module indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Quantity: 72 units

Why Was This Recalled?

Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Systems Corporation has 204 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report