Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

System 1 Base 220/240V The Terumo Advanced Perfusion System 1 Recalled by Terumo Cardiovascular Systems Corporation Due to Terumo Cardiovascular Systems (Terumo CVS) has received reports...

Date: June 25, 2012
Company: Terumo Cardiovascular Systems Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Terumo Cardiovascular Systems Corporation directly.

Affected Products

System 1 Base 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Quantity: 648 units

Why Was This Recalled?

Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. Occlusion malfunctions include stiff/difficult to adjust occlusion, loss of occlusion setting, unequal roller-to-roller occlusion and inability to adjust the occlusion. If it is

Where Was This Sold?

This product was distributed nationwide across all 50 states.