Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Serial interface module RS-232 Product Usage: The System 1 provides Recalled by Terumo Cardiovascular Systems Corporation Due to Terumo Cardiovascular Systems has received reports of malfunctions...

Date: July 26, 2012
Company: Terumo Cardiovascular Systems Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Terumo Cardiovascular Systems Corporation directly.

Affected Products

Serial interface module RS-232 Product Usage: The System 1 provides perfusion case data transfer through the use of the data transfer module. This module is a general purpose serial interface and is the interface between the TLink Data Management ystem or Sarns CCPro Data Perfusion oftware and the system. The data collected and exported by the data transfer module includes physiological data and device events to record the progress and events during a procedure.

Quantity: 579 units

Why Was This Recalled?

Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Systems Corporation has 204 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report