Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Custom Cardiovascular Procedure Kits Recalled by Terumo Cardiovascular Systems Corp Due to The Overpressure Safety Valve is not cleared with...

Date: May 6, 2015
Company: Terumo Cardiovascular Systems Corp
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Terumo Cardiovascular Systems Corp directly.

Affected Products

Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed it.

Quantity: 163 devices

Why Was This Recalled?

The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal blood loss, fluid weeping, or dripping on the user.

Where Was This Sold?

This product was distributed to 2 states: FL, MO

Affected (2 states)Not affected

About Terumo Cardiovascular Systems Corp

Terumo Cardiovascular Systems Corp has 19 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report