Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit Recalled by Teleflex Medical Due to The safety cap attached to needles within the...

Date: October 8, 2019
Company: Teleflex Medical
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Teleflex Medical directly.

Affected Products

Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Quantity: 526,000 total

Why Was This Recalled?

The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.

Where Was This Sold?

This product was distributed nationwide across all 50 states.