Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Teleflex HORIZON Microclip Titanium Ligating Clips w/tape Recalled by TELEFLEX LLC Due to Teleflex LLC is initiating this voluntary field corrective...

Date: April 28, 2023
Company: TELEFLEX LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact TELEFLEX LLC directly.

Affected Products

Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation

Quantity: 146,606 units

Why Was This Recalled?

Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures.

Where Was This Sold?

worldwide