Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tecan Freedom EVO with Touch Tools Suite Version 3.0 The Recalled by Tecan US, Inc. Due to WIth TouchTools 3.0 after the execution of a...

Date: February 19, 2013
Company: Tecan US, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Tecan US, Inc. directly.

Affected Products

Tecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run previously defined scripts and processes on a Freedom EVO platform (Touch Tools Suite is an optional software add-on for Freedom EVOware (the base operational software for the Freedom EVO 2 instrument.

Quantity: 12 units

Why Was This Recalled?

WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without warning). This could lead to the instrument moving unexpectedly and potentially cause an injury to the operator if they are physically inside t

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Tecan US, Inc.

Tecan US, Inc. has 19 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report