Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
TISSUE APPROXIMATION SYSTEM (TAS) Recalled by TAS Medical Inc Due to Tissue approximation straps (zip-ties) have broken before or...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact TAS Medical Inc directly.
Affected Products
TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Quantity: 114
Why Was This Recalled?
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About TAS Medical Inc
TAS Medical Inc has 1 total recall tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report