Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.
Battery Operated infusion pumps. t:slim X2 G5 Recalled by Tandem Diabetes Care, Inc. Due to Insulin pumps manufactured with speakers from Revision A...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Tandem Diabetes Care, Inc. directly.
Affected Products
Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement, REF: 1006379; Pump, t:slim X2, Control-IQ 7.4, REF: 1006402; Pump, t:slim X2, Control-IQ 7.4, Replacement, REF: 1006406; Pump, t:slim X2, Control-IQ 7.4, REF: 1006408; Pump, t:slim X2, Basal-IQ, , REF: 1006419; Pump, t:slim X2, Basal-IQ, mg/dL, REF: 1006420; Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement, REF: 1006535; Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement, REF: 1006537; Pump, t:slim X2, Control-IQ, mg/dl , Replacement, REF: 1008255; Pump, t:slim X2, Control-IQ, mmoUL, Replacement, REF: 1008256; Pump, t:slim X2, Basal-IQ, mmoUL, Replacement, REF: 1008257; Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement, REF: 1008258; Pump, t:slim X2, Basal-IQ, REF: 1010004; Pump, t:slim X2, Basal-IQ, Replacement, REF: 1010005; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010006; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010007; Pump, t:slim X2, Control-IQ, REF: 1010009; Pump, t:slim X2, Control-IQ Replacement, REF: 1010010; Pump, t:slim X2, Control-IQ, Medicare, REF: 1010011; Pump, t:slim X2, Control-IQ, Refurbished Replacement, REF: 1010012; Pump, t:slim X2, Control-IQ, REF: 1012867; Pump, t:slim X2, Control-IQ, Medicare, REF: 1012868; Pump, t:slim X2, Control-IQ, Replacement, REF: 1012869; Pump, t:slim X2, Control-IQ 7.8, REF: 1013195; Pump, t:slim X2, Control-IQ 7.8, Medicare, REF: 1013196; Pump, t :slim X2, Control-IQ 7.8, Replacement, REF: 1013197; Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement, REF: 1013198
Quantity: 171,750
Why Was This Recalled?
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Tandem Diabetes Care, Inc.
Tandem Diabetes Care, Inc. has 7 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report