Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Sysmex XE-2100C Automated Hematology System Recalled by Sysmex America, Inc. Due to There is a variation of reticulocyte counts between...

Date: June 1, 2012
Company: Sysmex America, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sysmex America, Inc. directly.

Affected Products

Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

Quantity: 69 units

Why Was This Recalled?

There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected