Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6. Recalled by Sysmex America, Inc. Due to Sysmex was made aware of a report of...

Date: January 6, 2025
Company: Sysmex America, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sysmex America, Inc. directly.

Affected Products

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.

Quantity: 32 units

Why Was This Recalled?

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Sysmex America, Inc.

Sysmex America, Inc. has 11 total recalls tracked by RecallDetector.

Related Recalls