Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets Recalled by Sysmex America, Inc. Due to There is potential that fluorescent platelet (PLT-F) and...

Date: September 8, 2017
Company: Sysmex America, Inc.
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sysmex America, Inc. directly.

Affected Products

Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers

Quantity: 23,239 units (9353 US)

Why Was This Recalled?

There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.

Where Was This Sold?

This product was distributed nationwide across all 50 states.