Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Recalled by Synthes (USA) Products LLC Due to TFNA nails may have the locking mechanism too...

Date: November 2, 2015
Company: Synthes (USA) Products LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Synthes (USA) Products LLC directly.

Affected Products

Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System) The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures. Intramedullary fixation rod, orthopedic.

Quantity: 5

Why Was This Recalled?

TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the bottom of the cannulated connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.

Where Was This Sold?

This product was distributed to 4 states: CA, FL, MI, WI

Affected (4 states)Not affected

About Synthes (USA) Products LLC

Synthes (USA) Products LLC has 162 total recalls tracked by RecallDetector.

Related Recalls