Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm Recalled by Synthes (USA) Products LLC Due to It was reported that the specified lots of...

Date: June 28, 2016
Company: Synthes (USA) Products LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Synthes (USA) Products LLC directly.

Affected Products

Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone, Orthopedic Devices The 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm is a part of the following systems: - 2.4. mm Titanium LCP Distal Radius Plate System - Titanium LCP Compact Distal Radius System - 2.4 mm Titanium LCP Radial Head Plate Set - Titanium Modular Mini Fragment LCP System - 2.4 mm Titanium Variable Angle LCP Distal Radius System - 2.7 mm/3.5 mm Titanium Distal Fibula Plate System

Quantity: 41 units

Why Was This Recalled?

It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.

Where Was This Sold?

This product was distributed to 17 states: AR, CA, CO, DE, FL, GA, IL, IA, KY, LA, MN, NJ, OK, PA, SD, TN, WA

Affected (17 states)Not affected

About Synthes (USA) Products LLC

Synthes (USA) Products LLC has 162 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report