Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

TFN-ADVANCED Proximal Femoral Nailing System (TFNA) Recalled by Synthes, Inc. Due to TFNA nails from certain lots were assembled with...

Date: April 9, 2015
Company: Synthes, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Synthes, Inc. directly.

Affected Products

TFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.

Quantity: 8175

Why Was This Recalled?

TFNA nails from certain lots were assembled with a locking mechanism too close to the top of the nail. Potential patient impact: surgical delay.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected