Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

Sunquest Laboratory : intended for use by professionals working in Recalled by Sunquest Information Systems, Inc. Due to Online calculations may not be performed as expected...

Date: November 15, 2011
Company: Sunquest Information Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sunquest Information Systems, Inc. directly.

Affected Products

Sunquest Laboratory : intended for use by professionals working in a clinical laboratory.

Quantity: 127 sites

Why Was This Recalled?

Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Sunquest Information Systems, Inc.

Sunquest Information Systems, Inc. has 9 total recalls tracked by RecallDetector.