Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Target Detachable Coils: TARGET HELICAL NANO 1 MM X 1 Recalled by Stryker Neurovascular Due to Stryker Neurovascular has become aware that some units...

Date: June 6, 2016
Company: Stryker Neurovascular
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Neurovascular directly.

Affected Products

Target Detachable Coils: TARGET HELICAL NANO 1 MM X 1 CM MODEL Number:M0035431010 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Quantity: 45,302 (8,188 in the U.S.) - total, all model numbers

Why Was This Recalled?

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Neurovascular

Stryker Neurovascular has 108 total recalls tracked by RecallDetector.