Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Stryker Neurovascular Recalled by Stryker Neurovascular Due to A design related issue can contribute to the...

Date: April 18, 2019
Company: Stryker Neurovascular
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Neurovascular directly.

Affected Products

Stryker Neurovascular, AXS Infinity LS Plus Long Sheath, Rx Only, Sterile EO, USA Customer Service 855-91 NEURO (916-3876): a) REF: INC-11196-70, 70cm, b) REF: INC-11196-80, 80cm, c) REF: INC-11196-90, 90cm Product Usage: Indications for Use: The AXS Infinity LS Plus Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. Device Description: The AXS Infinity LS Plus Long Sheath is a single lumen, flexible, variable stiffness long sheath with an 0.091 inch inner diameter. It has a radiopaque marker band on the distal end and a Luer hub at the proximal end. The AXS Infinity LS Plus Long Sheath shaft has a 10 cm lubricious coating at the distal end to reduce friction during use. The inner lumen of the AXS Infinity LS Plus Long Sheath is compatible with 6F or smaller catheters. Each package includes one AXS Infinity LS Plus Long Sheath (INC- 11196-70, INC-11196-80, or INC-11196-90), one Dilator, and one hemostasis valve. Dimensions of the AXS Infinity LS Plus Long Sheath are included on the individual device label.

Quantity: 2435

Why Was This Recalled?

A design related issue can contribute to the device exhibiting brittle behavior. This behavior may result in fractures of the catheter shaft.

Where Was This Sold?

This product was distributed nationwide across all 50 states.