Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Power-PRO 2 . Intended as an ambulance cot to transport Recalled by Stryker Medical Division of Stryker Corporation Due to Transport Cot has insufficient volume of epoxy to...

Date: February 1, 2023
Company: Stryker Medical Division of Stryker Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Medical Division of Stryker Corporation directly.

Affected Products

Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Model: 650705550001

Quantity: 4 units

Why Was This Recalled?

Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Medical Division of Stryker Corporation

Stryker Medical Division of Stryker Corporation has 79 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report