Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medi-Therm Hyper/Hypothermia System Recalled by Stryker Medical Division of Stryker Corporation Due to The water may reheat too quickly if the...

Date: March 29, 2017
Company: Stryker Medical Division of Stryker Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Medical Division of Stryker Corporation directly.

Affected Products

Medi-Therm Hyper/Hypothermia System, MTA7900

Quantity: 119

Why Was This Recalled?

The water may reheat too quickly if the unit is in "Automatic" mode, switched to "Manual" mode to view water temperature, and then switched back to "Automatic" mode.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Medical Division of Stryker Corporation

Stryker Medical Division of Stryker Corporation has 79 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report