Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: Recalled by Stryker Medical Division of Stryker Corporation Due to It was identified that the Mistral-Air Premium blankets...

Date: July 21, 2017
Company: Stryker Medical Division of Stryker Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Medical Division of Stryker Corporation directly.

Affected Products

MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM

Quantity: 11160

Why Was This Recalled?

It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).

Where Was This Sold?

U.S.

About Stryker Medical Division of Stryker Corporation

Stryker Medical Division of Stryker Corporation has 79 total recalls tracked by RecallDetector.