Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SurgiCount+ System Application Recalled by Stryker Instruments Div. of Stryker Corporation Due to When scanning sponges out after a surgical procedure,...

Date: June 15, 2022
Company: Stryker Instruments Div. of Stryker Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Instruments Div. of Stryker Corporation directly.

Affected Products

SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting, displaying, recording and locating the number of RFID-tagged absorbent articles used during surgical procedures.

Quantity: 44 units

Why Was This Recalled?

When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the occurrence of this error.

Where Was This Sold?

This product was distributed to 2 states: KS, ME

Affected (2 states)Not affected

About Stryker Instruments Div. of Stryker Corporation

Stryker Instruments Div. of Stryker Corporation has 161 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report