Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Stryker Instruments0400-750-000 T4 Toga Recalled by Stryker Instruments Div. of Stryker Corporation Due to Based on an accelerated aging test for this...

Date: May 23, 2013
Company: Stryker Instruments Div. of Stryker Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Instruments Div. of Stryker Corporation directly.

Affected Products

Stryker Instruments0400-750-000 T4 Toga, Pullover, Regular Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Quantity: 3,000 units (300 boxes)

Why Was This Recalled?

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Instruments Div. of Stryker Corporation

Stryker Instruments Div. of Stryker Corporation has 161 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report