Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Neptune 2 Waste Management System Docking Station (120V and 230V) Recalled by Stryker Instruments Div. of Stryker Corporation Due to The Fluid Coupling Sleeve, a component within the...

Date: July 20, 2015
Company: Stryker Instruments Div. of Stryker Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Instruments Div. of Stryker Corporation directly.

Affected Products

Neptune 2 Waste Management System Docking Station (120V and 230V), New and Refurbished Product Usage: The Stryker Docking Station (Docker) is a component of the Neptune 2 Waste Management System. The Stryker Rover, another component of the system, is a mobile unit used to suction and collect fluid waste and surgical smoke from a surgical site in an operating room. After collection, the Docker is mated to the Rover and the emptying of fluid waste and cleaning of the canisters occurs automatically. The Actuator Assembly is a component located within the Docker which consists of the fluid coupling parts under a spring loaded cover. Intended Use: The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor s offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electrocautery or laser devices.

Quantity: 524, 237 of these are non-conforming

Why Was This Recalled?

The Fluid Coupling Sleeve, a component within the Neptune Waste Management System Docker, may be out of specification. This can cut or dislodge the PTFE seal and cause a leak between the Rover and Docker couplers during docking or leaking from the Rover coupler after docking. The potential hazard is waste fluid and/or fresh water dispensed from the Neptune 2 Rover at the incorrect time.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Instruments Div. of Stryker Corporation

Stryker Instruments Div. of Stryker Corporation has 161 total recalls tracked by RecallDetector.