Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
C2 Nerve Monitor Recalled by Stryker Instruments Div. of Stryker Corporation Due to Stryker C2 NerveMonitor's design and instructions may not...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Instruments Div. of Stryker Corporation directly.
Affected Products
C2 Nerve Monitor, 4 Channel, 5140-508-240, UDI 07613327177329 Product Usage: The C2 Nerve Monitor System is intended for intra-operative monitoring and stimulation for localization and identification of cranial and peripheral motor and mixed motor sensory nerves during surgery.
Quantity: 135
Why Was This Recalled?
Stryker C2 NerveMonitor's design and instructions may not optimally address usability issues related to the functionality of the device, which may result in use errors potentially causing or contributing to nerve injury.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Stryker Instruments Div. of Stryker Corporation
Stryker Instruments Div. of Stryker Corporation has 161 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026